Validation Department
Validation
Overview
Validation is applied to many aspects of pharmaceutical manufacturing,
including instrumentation, HVAC systems etc. In each case the objective
is to produce “documented evidence, which provides a high
degree of assurance that all parts of a facility will consistently
work correctly when brought on-line”. Traditionally, validation
has consisted of the following:
Design Qualification (DQ) -
Documented evidence that the design of the system meets requirements
Installation Qualification (IQ) - System has been installed correctly
Operational Qualification (OQ) - System operates according to design
Performance Qualification (PQ) - System meets design criteria, and
operates according to requirement
Traditionally, each of the ‘Qualification’
sections has contained both the specification and the test protocols
aimed at verifying that each stage is completed. The term Validation
Protocol is often used when referring to these documents.
SCS Provide Documents which
simplify the overall validation process. Many of the tests performed
by SCS meet (or exceed) the site requirements for IQ and OQ, and
often become reference documents for the User to direct a testing
matrix toward.
System Documentation
The document which initiates the validation process is the User
Requirement Specification (URS). This describes the equipment or
system as it is supposed to work and is normally written by the
system user. The original version issued for quotation should normally
contain the essential requirements (musts) and the desirable requirements
(wants). The final version then accommodates all the musts, which
can be met, and any of the wants, which also can be satisfied. The
URS is now a standard document required for companies to meet the
ISO9001: 2000 standards.
The Function Design Specification
(FDS or FS) is written by SCS and describes the functions of the
equipment or system. The FDS links to the OQ as each major parameter
stated should be tested. The FDS is also a standard output document
required by ISO 9001: 2000.
The Hardware Design Specification
(HDS) is written by SCS and describes system architecture and hardware
design requirements.
The Software Design Specification
(SDS) is written by SCS and describes all data, architectural, interface
and component-level design for the software.
The documentation then includes
the system build and testing of the installed components. These
documents are the testing qualifications IQ, OQ and PQ. At given
stages during the lifecycle of the project, planned and systematic
reviews should be performed. The design reviews evaluates the deliverables
against standards and requirements.
Benefits of Validation
Validation is the formal documentation of good system development
practices. The principal benefit of validating a system is Compliance.
The information gathered before, during and after validation leads
to improved and confirmed compliance as changes are introduced.
Other benefits are:
• Systems which are well defined and specified, easier to
maintain, resulting in less downtime
• The system delivered is fit for purpose, on time, within
budget
• Satisfy the User Requirements
• In-depth understanding of the process
• Improved operational efficiency in a shorter time frame
• Reduced risk of failure
• Maintenance of quality standards
• Operators understanding the systems, leading to improved
communications between departments
Maintaining the Validated State
When a system is validated and in operation, measures should be
taken to ensure that the system remains in a validated state. This
maintenance not only involves the integrity of the hardware and
software, but also the documentation. The maintenance of a validated
system includes many activities, which the User is responsible for.
System Operational Procedures
(SOPs)
SOPs should be established to define the use and support of the
automated system. SOPs should be approved prior to use.
Training
Training plans should be established for use and support of the
system. These plans need to consider the training of all users,
technical and support.
Validation Engineers
As a Validation Engineer for
Standard Control Systems Paul Ryan produces documented evidence
of the Validated Building Management System which includes the Functional
specification (FS), Hardware Design Specification (HDS) & Software
Design Specification (SDS) and the Installation Qualification Test
(IQ) & Operation Qualification Test (OQ) which meet (or exceed)
the requirement of the User Requirement Specification (URS) and
compliant with FDA 21 CFR part 11.
Paul defines the Validated Building
Management System which leads to easier maintenance, resulting in
less downtime
E nsures the Validated Building Management System delivered is fit
for purpose, on time, within budget
Oversees the design of the Graphical User Interface (GUI)
P rovides in-depth training to the end user on the Graphical User
Interface
Paul has been involved with
implementing a Validated Building Management Systems on the following
pharmaceutical companies:
Allergan Pharmaceuticals (Ireland) Ltd
Servier (Ireland) Industries Ltd
Wyeth Medica Ireland
You can contact us at
Info@Standardcontrol.ie
Tel:01 4291800